Wednesday, 30 March 2011

Trace N2 Gas Analyser

Trace Nitrogen Gas Analysers

About:
The AGC Series 100 SED is engineered for the continuous monitoring of N2 in Ar or N2 in He.
Its N2 specific detector has been derived from our extremely successful High Frequency Argon Discharge Detector (ADD) system, the Series 100 HF-ADD Gas Analyser.

AGC Instruments has been providing this system configuration for several years to the gas manufacturing companies and especially to argon and helium specific plants throughout the world.

Principal:
The emission wavelength intensity of the Nitrogen molecule is selected to determine the trace amount of Nitrogen in Argon or Helium Matrix Gas. The response is proportional to the nitrogen concentration in low ppb to high ppm range.

Technique:
Using a spectral emission technique, a unique quartz ionisation chamber is utilised in which the sample gas is ionised by a high frequency tuned oscillator with the resultant emission wavelength intensity of the nitrogen
component is directed through a narrow band pass filter and is detected via a photomultiplier tube.


http://www.agc-instruments.com/trace-nsub2sub-gas-analysers